McClellan Report - Sampling Devices
     

 
 geoinsightonline@yahoo.com
 phone 800-996-2225


 HYDRASLEEVE REFERENCE
 DOCUMENTS
Overview of Passive Sampler TechnologiesITRC Document Overview of Passive Sampler Technologies concludes HydraSleeve is an inexpensive, simple tool to collect groundwater samples without purging.
View the PDF
Passive Sampler Tech Matrix

Read the McClellan Air Force Base Final Report!McClellan Air Force Base Final Report rates HydraSleeve highest among no-purge groundwater sampling devices tested at the former McClellan AFB, California.
View the Report
Passive Samplers Comparison


 

Report Cover   Table of Contents   Sec. 1   Sec. 2   Sec. 3   Sec. 4   Sec. 5   Sec. 6   Sec. 7   Sec. 8

Download the Report PDF (874 KB)
("Right-click", then "Save as..." to your desktop)
View Report Appendices

SECTION 3

FIELD ACTIVITIES AND LABORATORY ANALYTICAL APPROACH

3.1 FIELD ACTIVITIES

A total of 251 primary samples and 34 quality assurance/quality control (QA/QC) samples were collected from 20 wells at McClellan as part of this demonstration. Details of the field activities are discussed below.

3.1.1 SAMPLING STRATEGY

Concurrent deployment of multiple types of samplers at the same depth in each well is desirable to obtain comparative data. However, the 4-inch well diameter imposed a physical limitation on the number of samplers that could be concurrently deployed at the same depth in each well. Therefore, sampling occurred in three phases as described below.

  • Phase 1 – During this phase, which occurred from May 17 through 21, 2004, the diffusion samplers (PDBS, RPPS, PsMS, and RCS) were deployed in the 20 selected monitoring wells at three different depths per well. No more than 3 different types of diffusion samplers were deployed in each well.
  • Phase 2 – After an approximate 3-week equilibration period, the diffusion samplers deployed in Phase 1 were retrieved (from June 7 through 9, 2004). The grab samplers (Snap Sampler™ and HydraSleeve®) were subsequently deployed at the same depths as the samplers deployed in Phase 1. Only one type of grab sampler was deployed in each well; concurrent deployment of both the Snap Sampler™ and HydraSleeve® in the same 4-inch well would have made deployment and retrieval difficult and may have compromised the function of one or both of the devices..
  • Phase 3 – After an approximate 1-week equilibration period, the grab samplers were retrieved (from June 14 through 17, 2004). Following this retrieval, conventional sampling (i.e., low-flow purge/sample and three-volume purge/sample) of all 20 wells was performed. Both low-flow purge/sample and three-volume purge/sample techniques were used at each well.

Table 3.1 is a summary of the types of sampling techniques that were used in each well.

TABLE 3.1
SAMPLING TECHNOLOGIES DEMONSTRATED IN EACH WELL
NO-PURGE SAMPLER DEMONSTRATION
McCLELLAN AFB, CALIFORNIA


3.1.2 FIELD MEASUREMENTS

The depth to water was measured in each well prior to deployment during Phase 1, prior to retrievals during Phase 2, and prior to conventional sampling during Phase 3. Additionally, the total well depth was measured prior to deployment during Phase 1. Target sampler deployment depths were calculated after measuring the depth to water and the total well depth at the beginning of Phase 1, taking into consideration the reported screened interval of the well. Of the three sampling depths monitored per well, the intermediate interval was generally defined as the center of the saturated screened interval, the shallow interval was generally defined as being approximately 1 foot below the top of the saturated screened interval, and the deep interval was generally defined as being approximately 1 foot above the bottom of the open (i.e., non-buried) saturated screened interval. Table 3.2 is a summary of the depth to water measurements, the total depth measurements, the screened interval depths, and the sampling intervals for each well.

TABLE 3.2
WATER LEVEL MEASUREMENTS, WELL DETAILS,
AND DEPLOYMENT DEPTHS
NO-PURGE SAMPLER DEMONSTRATION
McCLELLAN AFB, CALIFORNIA
View TABLE 3.2

Measurements of traditional well stabilization parameters were made during conventional sampling. These parameters included groundwater temperature, pH, conductivity, DO, ORP, and turbidity. These measurements along with the total volume purged, the time spent purging, and the average pump rate for each well are summarized in Table 3.3.

A maximum of three different types of diffusion samplers and one type of grab sampler were deployed in each well. The distribution of diffusion and grab samplers in each well was designed to facilitate inter-sampler comparisons while maintaining an overall deployment of RPPS in 20 wells; RCS, PsMS, HydraSleeve®, and Snap Samplers in 10 wells each; and PDBS in only those wells that were targeted for VOC analysis. Table 3.4 is a summary of the sample dates, deployment lengths, and time lags between all sampling events.

3.2 LABORATORY ANALYTICAL APPROACH

3.2.1 TARGET COMPOUNDS

The following compounds were targeted for analysis in the priority listed below during the technology demonstration.

  • 1,4 dioxane;
  • Hexavalent chromium;
  • McClellan target analyte list (TAL) for metals, total and/or dissolved phases depending on sample turbidity (see below and Section 4.2 of Work Plan [Parsons, 2004a]) including: aluminum, antimony, arsenic, barium, beryllium, cadmium, calcium, chromium, cobalt, copper, iron, lead, magnesium, manganese, molybdenum, nickel, potassium, selenium, silver, sodium, thallium, vanadium, and zinc;
  • Anions including sulfate, nitrate, and chloride; and
  • VOCs (refer to Table 4.11 of the McClellan QAPP [URS, 2003] for a list of specific analytes).

With the exception of VOCs, these compounds were targeted because they are not able to be monitored using the PDBS method, but are contaminants of concern at some DoD installations. VOCs were included in the target compound list to verify that all no-purge sampling devices also would be capable of accurately monitoring for these compounds.

The final measurements of turbidity made during both types of conventional sampling were used to determine whether or not the samples should be field-filtered for TAL metals analysis using a 0.45-micron disposable filter. If the final turbidity measurement made immediately before sample collection was less than or equal to 5 Nephelometric Turbidity Units (NTUs), the samples were not filtered in the field and were submitted for total metals analysis. If the final turbidity measurement was greater than 5 NTUs, the samples were filtered according to procedures described in SOP #6 of the Work Plan (Parsons, 2004a), and were scheduled for dissolved metals analysis. All conventionally sampled wells that were analyzed for metals were field-filtered with the exception of well

TABLE 3.3
SUMMARY OF CONVENTIONAL SAMPLING
FIELD PARAMETER MEASUREMENTS
NO-PURGE SAMPLER DEMONSTRATION
McCLELLAN AFB, CALIFORNIA
View TABLE 3.3

TABLE 3.4
SAMPLE DATES AND TIME LAGS
NO-PURGE SAMPLER DEMONSTRATION
McCLELLAN AFB, CALIFORNIA
View TABLE 3.4

MW-400 where the measured turbidity was less than 5 NTUs. Additionally, all metals samples collected using the HydraSleeve® were field-filtered. Samples for hexavalent chromium analysis were not field-filtered.

3.2.2 LABORATORIES

Two analytical laboratories were used during this demonstration to perform all of the required analyses. Columbia Analytical Services, Inc. (CAS) in Kelso, Washington performed the metals and 1,4 dioxane analyses. Sequoia Analytical (Sequoia), based in Sacramento, California performed the hexavalent chromium, anion, and VOC analyses. Sequoia used two different facilities to perform the requested analyses; hexavalent chromium and anions were analyzed in their Morgan Hill, California facility while VOCs were analyzed in their Petaluma, California facility.

The maximum holding time permitted for hexavalent chromium is 24 hours. Therefore, samples were sent twice per day (once at approximately noon, and again at approximately 5 pm) to Sequoia using a hand-delivery courier. Samples were shipped daily each afternoon to CAS via overnight express courier.

3.2.3 SAMPLE VOLUME

As described in the Work Plan (Parsons, 2004a), the diffusion and grab samplers do not collect large volumes of groundwater (relative to conventional sampling methods), and the available sample volume does not always fulfill normal laboratory and/or analytical method recommendations. This characteristic is not necessarily a critical limitation since most analytical methods do not actually require the larger sample volumes recommended in standard analytical procedures. An ITRC Diffusion Sampler subteam has estimated the minimum sample volumes required for common environmental analytical methods; details are available on the ITRC diffusion sampling website at http://64.203.146.40/news.asp#41. Prior coordination with the analytical laboratories enabled use of smaller sample volumes to perform the required analytical methods while still maintaining required detection limits. Table 3.5 is a summary of the approximate maximum volume capacities of each type of no-purge sampling device used in this study per sample depth (some sampling devices required more than one sampler per depth interval). The volumes listed in Table 3.5 are the maximum obtainable with the configuration used at McClellan; larger volumes can potentially be obtained in some cases by reconfiguring the samplers (e.g., using more PsMS canisters). It should be noted that a larger-volume Snap Sampler™ and HydraSleeve® are now available. Table 3.6 summarizes the minimum sample volume requirements (per analysis) specified by the analytical laboratories.

Groundwater samples from each well were analyzed for only a subset of the target analyte list. The minimum sample volumes shown in Table 3.6 were used for diffusion, grab, and low-flow samples to maintain consistency and to facilitate comparison of the results. However, in order to maintain consistency between the three-volume purge method historically used for these wells as part of LTM and the conventional samples collected as part of this demonstration, normal sample volumes specified in the McClellan QAPP (URS, 2003) were collected for the three-volume purge method.

TABLE 3.5
VOLUMETRIC CAPACITIES OF SAMPLING DEVICES
NO-PURGE SAMPLER DEMONSTRATION
MCCLELLAN AFB, CALIFORNIA


TABLE 3.6
MINIMUM VOLUME REQUIREMENTS
NO-PURGE SAMPLER DEMONSTRATION
MCCLELLAN AFB, CALIFORNIA


One or more additional sets of sample bottles were filled and submitted to the analytical laboratory along with the primary sample whenever sufficient sample volume was available. This practice allowed the laboratory to reanalyze samples as necessary due to the need for sample dilution or other circumstances.

3.3 DEVIATIONS FROM WORK PLAN

The field activities generally occurred in accordance with the Work Plan (Parsons, 2004a). However, the following notable deviations occurred during this evaluation.

  • While measuring the total depth of MW-1031, the depth sounding device continually became caught on the inside of the well. Due to concerns of having the no-purge samplers stuck or damaged inside the well during deployment and/or retrieval, MW-1031 was replaced with the first alternate well (MW-424) listed in the Work Plan (Parsons, 2004a).
  • Upon retrieval of the HydraSleeve® samplers from well MW-148, a knot was observed in the rope used for deployment approximately 10 to 30 feet above the top of the upper (i.e., shallow) sampler. Approximately 10.8 feet of rope was tangled as part of this knot, which presumably meant that all HydraSleeve® samplers in this well were actually deployed approximately 10.8 feet higher in the well than anticipated. Additionally, upon retrieval the deepest HydraSleeve® sampler from this well had a hole in it and no water was recovered. Because of these issues, only the intermediate depth sampler was sent to the laboratory for analysis.
  • The trigger mechanism of the Snap™ Sampler was not pulled hard enough at two wells, resulting in no Snap™ Samples being collected from well MW-242 and no deep Snap Sample being collected from well MW-427.
  • In order to evaluate the ability of PDBS to monitor 1,4 dioxane, this analysis was requested for one PDBS during the demonstration (the shallow PDBS deployed in well MW-72).
  • The measured total depth in well MW-38D was 120.8 ft bgs (Table 3.2). The reported values for the top and bottom of the screened interval for this well were 120.03 and 130.03 ft bgs, respectively. Based on these values, only approximately 0.8 foot of screen was open in this well. Accordingly, only one depth interval was monitored (defined as the deep interval) at 120.3 ft bgs.
  • The water level indicator used during the Phase 2 activities malfunctioned during the afternoon of June 8, 2004. Accordingly, no water level measurements were obtained for the last 1.5 days of Phase 2 activities.
  • Hexavalent chromium was analyzed using US Environmental Protection Agency (USEPA) Method SW7199 as opposed to SW7196M as described in the Work Plan (Parsons, 2004a). Use of SW7199 permitted a lower detection limit than would have been possible with SW7196M.
  • Metals were analyzed using USEPA Methods SW6010, SW6020, and SW7740 as opposed to only SW6020 as described in the Work Plan (Parsons, 2004a).
  • Typically, at least two 20-mL VOA vials were shipped to Sequoia for VOC analysis. The expectation (based on prior discussions with the laboratory) was that Sequoia would use one sample bottle for the initial analysis and would use any additional sample bottles as back-up samples in the event that re-analysis was necessary (e.g., dilutions). However, for most analyses Sequoia composited the two 20-mL VOA vials into one 40-mL VOA vial for analysis using their autosampler. Parsons discussed this issue with Sequoia after realizing that the procedure was being used, and Sequoia clarified that the procedure that was used was consistent with USEPA guidance. Nonetheless, the potential for volatilization of VOCs during the compositing process is a potential concern.
  • Due to a field oversight, hexavalent chromium was not analyzed in either the Low- Flow or the three-Volume samples collected from wells MW-38D and MW-424.

3.4 QA/QC SAMPLE COLLECTION

A total of 34 samples were collected for QA/QC purposes. The number and type of each of these samples is summarized in Table 3.7. Generally QA/QC sample collection followed the schedule described in the Work Plan (Parsons, 2004a). However, some variances did occur as described below.

Sequoia did not provide trip blank samples as part of the Phase 2 bottle order. However, one trip blank sample was provided via courier by Sequoia on June 9, 2004. This was the only trip blank sample collected during the Phase 2 activities. This sample was sent to Sequoia along with the daily shipment of VOC samples on June 9, 2004. However, Sequoia did not analyze this sample. No explanation was available from Sequoia as to why this sample was not analyzed. Trip blank samples were provided by Sequoia for the Phase 3 activities, and one of these samples was shipped along with each

TABLE 3.7
QA/QC SAMPLES COLLECTED
NO-PURGE SAMPLER DEMONSTRATION
McCLELLAN AFB, CALIFORNIA

a/ Although four samples were collected with the intention of being used as field duplicates, a fifth field duplicate sample was available for the analyses performed by Sequoia (see Note b/ below).

b/ These samples were designated for MS/MSD analyses on the chains of custody. However, Sequoia treated them as primary samples and did not spike them. They therefore are considered field duplicate samples for analyses performed by Sequoia only. Although no other samples were designated by the field scientists as MS/MSD samples, both Sequoia and CAS chose other samples at random upon which to perform MS/MSD analyses (see Appendix A).

c/ Source water blank was comprised of the water used to fill the diffusion samplers prior to deployment.

d/ NA = not applicable.

e/ Purified water blank was comprised of the water used for decontamination.

cooler containing samples intended for VOC analysis. As a result of the lack of trip blanks during Phase 2, the degree to which low-level VOC detections may be attributable to cross-contamination during sample shipping and handling cannot be fully confirmed. Two of the samples collected with the intent of being used by the laboratories as matrix spike/matrix spike duplicate (MS/MSD) samples were not treated as MS/MSD samples by Sequoia although they were by CAS. Instead, Sequoia analyzed these samples as primary samples. They are therefore considered duplicate samples for QA/QC purposes. These samples were MW173-3VOL-MS/MSD and MW225-MICROMS/ MSD. Despite this oversight, other samples were selected at random by Sequoia for MS/MSD analysis (see Appendix A). In the instances where field samples designated as MS/MSDs were not analyzed as such, measurements of accuracy and analytical precision based on MS/MSD results were not developed for samples collected using a given sampling method.

In the Work Plan (Parsons, 2004a), two field duplicates and two MS/MSD samples were scheduled for collection with the HydraSleeve®. However, due to an oversight, no field duplicates or MS/MSD samples for this sampler type were collected. Therefore, information regarding precision of the HydraSleeve® sampling process based on MS/MSD results and the impact of potential matrix effects on the analytical testing is not available.

A total of four field duplicate samples were collected for both the low-flow and threevolume purge sampling methods while only two were scheduled according to the Work Plan (Parsons, 2004a).

Although only one equipment rinseate was scheduled for the three-volume purge method (Parsons, 2004a), two were actually collected; one from the bailer only, and another from both the bailer and the in-line filter.

Report Cover   Table of Contents   Sec. 1   Sec. 2   Sec. 3   Sec. 4   Sec. 5   Sec. 6   Sec. 7   Sec. 8

Copyright © 2008  •  GeoInsight  •  800-996-2225
No-Purge Groundwater Sampling  •  Passive Ground Water Sampling